Operation
Continuous Operation
The Odyssey F is designed to operate continuously. During idle times, the Odyssey F can remain powered on or be powered off at your discretion. When idle, the Odyssey F typically draws approximately 0.5 Amp of current when powered with 120V (see Odyssey® F Imager Specifications). If the Odyssey F needs to be disconnected (when moving it, for example), see Disconnecting the Power.
Before You Begin...
It is critical for the scan surface glass to be free from smudges, dust, and dye before scanning any samples on the Odyssey F Imager. The silicone mat, slide holder, and plate alignment guide must also be thoroughly cleaned before each use.
Contaminated surfaces in contact with samples may cause blotches and streaks that cannot be removed with further washing.
Clean the Scan Surface
Thoroughly clean the glass scan surface and the metal frame with ultrapure water and a lint-free wipe.
Do not use scouring compounds or abrasive scouring pads. The glass can scratch, which can affect the scanned image.
Repeat the wash with 70% ethanol to remove any visible smears. Use methanol to remove any remaining residues.
Do not get ethanol or methanol on the outside of the imager.
Clean Silicone Mat (926-70003)
Clean the silicone mat each time it is used.
Rinse the mat under warm water. Gentle lab soap may be used, but the soap must be completely rinsed away before use. The silicone mat may also be rinsed with isopropanol if needed.
Air dry the mat.
Clean Alignment Guides (926-18972 and )
The plate alignment guide (926-18972) and slide holder ( ) can be cleaned with soap and warm water. Clean the guides away from the imager to ensure that the soapy water does not enter the imager.
Keeping these guides clean will help reduce wear over time and will help ensure proper connection between the permanent magnets in the guide with the magnets they connect to on the imager.
Disconnecting the Power
First, exit the software.
Image Studio Software: Select Instrument > Disconnect on the Application menu or by closing the application window.
LI‑COR Acquisition Software: Exit by closing the application window.
Briefly press the power button on the front panel to power off the Odyssey F Imager. Then remove the power cord from the power receptacle as described in Safety.
Note: Do not unplug the power cable while the system is running, except for an emergency situation.