Odyssey® F Imager Specifications

Detection Channel Summary

ModelChannelExcitationEmission
9150, 9180800785 nm813 – 832 nm
9150, 9180700685 nm710 – 730 nm
9150, 9180700A685 nm813 – 832 nm
9180520520 nm570 – 610 nm
9180520A520 nm710 – 730 nm
9180520B520 nm813 – 832 nm
9180488488 nm538 – 552 nm
9180488A488 nm570 – 610 nm
9180488B488 nm710 – 730 nm
9180488C488 nm813 – 832 nm

Specifications

Unless otherwise specified, the following apply to both Odyssey F 9150 and Odyssey F 9180.

Imaging System

  • Dynamic Range: >6 logs

  • Pixel Resolution: 21 μm, 42 μm, 84 μm, 169 μm, 337 μm

  • Focusing Range: -1.00 – 5.00 mm

  • Solid-state diode lasers:

    • Model 9150: 685, 785

    • Model 9180: 488, 520, 685, and 785

  • Detection Modes: Fluorescence

  • Laser lifetime: 40,000 hours of operation

  • Microplate Insert: Accessory alignment guide for 4 microplates

  • Microscope Slide Holder: Accessory holder for 12 slides

  • Image Scan Times

    • Western blot: 4 minutes for a 10 cm x 10 cm membrane scanned at 169 μm using 2 lasers at a time.

    • Microplate: 3 minutes for a single, 96-well microplate scanned at 169 μm using 2 lasers at a time.

  • Dimensions: 37 H × 53 W × 62 D cm (15 H × 21 W × 24.5 D inches) With the hood fully open, the Odyssey F requires 74 cm of vertical clearance.

  • Image field size: 25 cm W × 18 cm D (9.8” W × 7.1” D)

  • Weight: 32 kg (70 lbs)

Laser Source

The Odyssey F Imager is a class 1 laser product.

All lasers run at a 50% duty cycle between 20KHz and 50KHz and diverge at 6 degrees.

Laser SourceMaximum Internal Laser Power
488 nm12mW
520 nm20mW
685 nm20mW
785 nm50mW

Operating Specifications

Operating Conditions: For indoor use only; operating temperature 15 - 30 °C and dew point not greater than 22 °C, non-condensing; altitude not to exceed 2000 m.

Environmental Conditions: Pollution Degree 2.

Power Requirements: Universal input range is between 100-240 VAC (voltage fluctuations not to exceed 10% of the nominal voltage); 2.2 Amp maximum; at 120V, 0.5 Amp typical when idle; 50/60 Hz.

Network Protocol: TCP/IP.

Network Connection: Cat. 5e RJ45, 10BASE-T/100BASE-TX/1000BASE-T. Use only the supplied cable.