Operating Temperature: 0 to 50 °C with no solar load (non-freezing)
Storage Temperature: -20 to 60 °C with chamber clean and dry
Temperature Control: User provided water bath. #10-32 threaded connections to chamber.
Operating Fluid Environment:
Temperature: non-freezing to 50 °C
Salinity: 0 – 35 %
Auxiliary Ports:
pH (probe not included): 12 mm diameter O-ring sealed port and integrated amplifier. Passive glass-electrode based pH probe, with BNC connector (nominal -59 mV/pH slope, user calibrated).
Septa: Silicone-PTFE septa
Clear-top Chamber
Maximum Leaf Area: 9 cm2
Size: 15.4 x 11.5 x 5.9 cm (L x W x H)
Weight: 0.3 kg
Small Light Source
Total Output Range: 0 – >2000 µmol m-2s-1 at 25 °C
Blue Output Range: 0 – >400 µmol m-2s-1 at 25 °C
Red Output Range: 0 – >1600 µmol m-2s-1 at 25 °C
Red Peak Wavelength: 660 nm
Blue Peak Wavelength: 453 nm
Uniformity:
±10% Over 90% of the aperture with white top gasket, typically
±10% Over 77% of the aperture with black gasket, typically
Power Consumption: 2000 µmol m-2s-1 <5 watts
Operating Temperature Range: 0 – 50 °C
Size: 6.6 x 5.9 x 5.8 cm (L x W x H)
Weight: 0.21 kg
Large Light Source
Total Output Range: 0 - >2500 µmol m-2s-1 at 25 °C
Blue Output Range:>2000 µmol m-2s-1 at 25 °C
Green Output Range: >1000 µmol m-2s-1 at 25 °C
Red Output Range: >2400 µmol m-2s-1 at 25 °C
White Output Range: >1500 µmol m-2s-1 at 25 °C
Blue Peak Wavelength: 453 nm
Green Peak Wavelength: 523 nm
Red Peak Wavelength: 660 nm
White Color Temperature: 4000 K
Uniformity: ±10% over 90% of the aperture
Power Consumption: 15 W at 2000 µmol m-2s-1 equal parts red, green, blue, and white
Operating Temperature Range: 0 – 50 °C
Operating Relative Humidity Range: 0 – 85%
Size: 11.7 × 11 × 13 cm (L × W × H)
Weight: 0.54 kg
Small Plant Chamber
Chamber Volume: 193.2 cm3 (internal volume)
Internal Dimensions: 7 cm diameter; 4.46 cm deep
External Dimensions: 8.4 cm wide (chamber block) × 12.7 cm long (chamber manifold interface to tip of chamber) × 6.47 cm (bottom of chamber manifold to top of chamber)
Weight: 0.60 kg
Custom Chamber Adapter
Chamber Volume: 34.2 cm3 (internal volume)
External Dimensions: 3.18 × 7.67 × 5.85 cm (L × W × H)
Insect Respiration Chamber
Chamber Volume: 49.9 cm3 (not including tubing)
Exterior Dimensions: 11.25 cm length; 3 cm diameter
Weight: 0.07 kg
Multiphase Flash™ Fluorometer
Modulated Light: Software controlled and software selectable frequencies of 1 Hz – 250 kHz
Measuring Light Peak Wavelength: 625 nm
Red Actinic and Saturating Flash Peak Wavelength: 625 nm
Blue Actinic Peak Wavelength: 475 nm
Far-red Peak Wavelength: 735 nm
Actinic Light Output:
0 – 3000 µmol m-2s-1 total at 25 °C
0 – 1000 µmol m-2s-1 blue at 25 °C
0 – 2000 µmol m-2s-1 red at 25 °C
Saturation Light: Software controlled intensity; 0 – 16,000 µmol m-2s-1 at 25 °C
Far-red Light: Software controlled intensity; 0 – 20 µmol m-2s-1 at 25 °C
Fluorescence Signal Temperature Dependence: -0.25% per °C
Uniformity:
<±10% over 92% of the aperture with white top gasket
<±10% over 90% of the aperture with black top gasket
Power Consumption:
<18 W at 25 °C with 3000 µmol m-2s-1 actinic light
<60 W at 25 °C with 16,000 µmol m-2s-1 saturating flash
Leaf Area: 6 cm2 or 2 cm2; Round apertures
Size: 16.6 x 11.5 x 13.6 cm (L x W x H)
Weight: 0.86 kg
Large Leaf and Needle Chamber
Maximum Leaf Area: 36 cm2
External Dimensions: 11.5 cm wide (spring to spring) x 16.8 cm long (front of chamber to back of chamber link) x 5.9 cm tall (bottom of chamber manifold to top of chamber)
Weight: 0.35 kg
Large Leaf and Needle Chamber with Conifer Sprig Blocks
Maximum Planar Leaf Area: 36 cm2
Internal Chamber Height: 6.7 cm
External Dimensions: 11.5 cm wide (spring to spring) x 16.8 cm long (front of chamber to back of chamber link) x 7.2 cm tall (bottom to top of chamber)
Volume: 25.68 in3 or 420.8 cm3
Bryophyte Chamber
Chamber Volume: 193.2 cm3 (internal volume)
Internal Dimensions: 7 cm diameter; 4.45 cm deep
External Dimensions: 8.4 cm wide (chamber block) × 12.7 cm long (chamber manifold interface to tip of chamber) × 6.47 cm (bottom of chamber manifold to top of chamber)
Weight: 0.60 kg
Soil CO2 Flux Chamber
Chamber Volume: 4244.1 cm3
IRGA Volume: 57 cm3
Soil Area: 317.8 cm2 (49.3 in2)
Air Temperature Thermistor:
Operating Range: -20 to 45 °C
Accuracy: ±0.5 °C over 0 °C to 70 °C
Weight: 4.06 kg
SATP is defined as Standard Ambient Temperature (25 °C) and Pressure (100 kPa).