CO2 measurements
- Measurement range: 0 to 20,000 ppm
- Accuracy
- LI-850 Within 1.5% of reading
- LI-830 Within 3% of reading
- Calibration drift
- Zero Drift1: <0.15 ppm / °C
- Span Drift2: <0.03% / °C
- Total drift at 370 ppm: <0.4 ppm / °C
- RMS noise at 370 ppm with 1 sec signal filtering: <1 ppm
- Sensitivity to water vapor (LI-850 only): <0.1 ppm CO2 / mmol mol-1 H2O
- Lower limit of detection 1.5 ppm
H2O measurements (LI-850 only)
- Measurement range: 0 to 60 mmol mol-1
- Accuracy: Better than 1.5% of reading
- Calibration drift
- Drift at 0 mmol mol-1: <0.005 mmol mol-1 / °C
- Span Drift1 at 10 mmol mol-1: <0.006 mmol mol-1 / °C
- Total Drift at 10 mmol mol-1 <0.016 mmol mol-1 / °C
- RMS noise at 10 mmol mol with 1 sec signal filtering: <0.01 mmol mol-1
- Sensitivity to CO2: <0.0001 mmol mol-1 H2O / ppm CO2
- Lower limit of detection: <0.015 mmol mol-1
General
- Output rate: Up to 2 measurements per second
- Response time (T90)
- CO2: <3.5 seconds from 0 to 375 ppm
- H2O: <3.5 seconds from 0 to 21 mmol mol-1
- Internal optical cell volume: 14.5 mL
- Measurement principle: Non-Dispersive Infrared
- Traceability:
- CO2: Traceable gases to WMO standards from 0 to 3,000 ppm; traceable gases to EPA protocol gases from 3,000 to 20,000 ppm
- H2O (LI-850 only): NIST traceable LI-610 Portable Dew Point Generator
- Pressure compensation range: 50 to 110 kPa
- Maximum gas flow rate: 1 liter min-1
- Output signals: Two analog voltage (0 to 2.5 V or 0 to 5 V) and two current (4 to 20 mA)
- Digital outputs: TTL (0 to 5 V) or Open Collector
- DAC resolution: 16 bits across user-specified range
Power requirements
- Input voltage: 12-30 VDC
- After warmup (without pump): 0.33A @ 12 VDC (4.0 W) average
- After warmup (with pump): 0.42A @ 12 VDC (5.0 W) average
- During warmup: 1.2 A @ 12 VDC (14 W) maximum
- Operating temperature range: -20°C to 45 °C
- Relative humidity range: 0 to 95% RH, Non-condensing
- Dimensions: 22.23 cm W x 15.25 cm D x 7.62 cm H
Weight
- No pump, no display: 1.0 kg
- No pump, with display: 1.02 kg
- With pump, no display: 1.3 kg
- With pump, with display: 1.32 kg
Pump specifications (optional)
- Operating temperature range: 5 to 45 °C
- Storage temperature range: -20 to 60 °C
- Operating humidity range: 0 to 80% RH
- Nominal flow rate: 0.75 liters minute-1
- Power consumption: 1 W (nominally)
- Expected life span: 8,000 hrs in standard conditions with a normal load
Display specifications (optional)
- Dimensions: 6.7 cm corner-to-corner
- Resolution: 400 x 200 px; monochrome
- Power consumption: <200 µW
- Displayed variables: CO2 reading, H2O reading (LI-850 only), optical bench temperature, and pressure.
Specifications subject to change without notice.