Specifications

CO2 measurements

  • Measurement range: 0 to 20,000 ppm
  • Accuracy
    • LI-850 Within 1.5% of reading
    • LI-830 Within 3% of reading
  • Calibration drift
    • Zero Drift1: <0.15 ppm / °C
    • Span Drift2: <0.03% / °C
    • Total drift at 370 ppm: <0.4 ppm / °C
  • RMS noise at 370 ppm with 1 sec signal filtering: <1 ppm
  • Sensitivity to water vapor (LI-850 only): <0.1 ppm CO2 / mmol mol-1 H2O
  • Lower limit of detection 1.5 ppm

H2O measurements (LI-850 only)

  • Measurement range: 0 to 60 mmol mol-1
  • Accuracy: Better than 1.5% of reading
    • Calibration drift
    • Drift at 0 mmol mol-1: <0.005 mmol mol-1 / °C
    • Span Drift1 at 10 mmol mol-1: <0.006 mmol mol-1 / °C
    • Total Drift at 10 mmol mol-1 <0.016 mmol mol-1 / °C
    • RMS noise at 10 mmol mol with 1 sec signal filtering: <0.01 mmol mol-1
  • Sensitivity to CO2: <0.0001 mmol mol-1 H2O / ppm CO2
  • Lower limit of detection: <0.015 mmol mol-1

General

  • Output rate: Up to 2 measurements per second
  • Response time (T90)
    • CO2: <3.5 seconds from 0 to 375 ppm
    • H2O: <3.5 seconds from 0 to 21 mmol mol-1
  • Internal optical cell volume: 14.5 mL
  • Measurement principle: Non-Dispersive Infrared
  • Traceability:
    • CO2: Traceable gases to WMO standards from 0 to 3,000 ppm; traceable gases to EPA protocol gases from 3,000 to 20,000 ppm
    • H2O (LI-850 only): NIST traceable LI-610 Portable Dew Point Generator
  • Pressure compensation range: 50 to 110 kPa
  • Maximum gas flow rate: 1 liter min-1
  • Output signals: Two analog voltage (0 to 2.5 V or 0 to 5 V) and two current (4 to 20 mA)
  • Digital outputs: TTL (0 to 5 V) or Open Collector
  • DAC resolution: 16 bits across user-specified range

Power requirements

  • Input voltage: 12-30 VDC
    • After warmup (without pump): 0.33A @ 12 VDC (4.0 W) average
    • After warmup (with pump): 0.42A @ 12 VDC (5.0 W) average
    • During warmup: 1.2 A @ 12 VDC (14 W) maximum
  • Operating temperature range: -20°C to 45 °C
  • Relative humidity range: 0 to 95% RH, Non-condensing
  • Dimensions: 22.23 cm W x 15.25 cm D x 7.62 cm H

Weight

  • No pump, no display: 1.0 kg
  • No pump, with display: 1.02 kg
  • With pump, no display: 1.3 kg
  • With pump, with display: 1.32 kg

Pump specifications (optional)

  • Operating temperature range: 5 to 45 °C
  • Storage temperature range: -20 to 60 °C
  • Operating humidity range: 0 to 80% RH
  • Nominal flow rate: 0.75 liters minute-1
  • Power consumption: 1 W (nominally)
  • Expected life span: 8,000 hrs in standard conditions with a normal load

Display specifications (optional)

  • Dimensions: 6.7 cm corner-to-corner
  • Resolution: 400 x 200 px; monochrome
  • Power consumption: <200 µW
  • Displayed variables: CO2 reading, H2O reading (LI-850 only), optical bench temperature, and pressure.

Specifications subject to change without notice.