Factory Calibration
Recommended Recalibration Frequency: No regular factory recalibration required.
What is included with Trace Gas Analyzer recalibration?
The LI-7810, LI-7815, LI-7820, or LI-7825 factory recalibration includes the following adjustments and tests:
Calibration
- Zero, span, and verification.
- Pressure sensor calibration.
Tests and Maintenance
- Plumbing inspection and leak test.
- Replace dry-down chemical.
- Change user-replaceable inlet filters.
- Inspect O-rings and filter housing on air inlet and outlet.
- Lithium coin cell battery changed (as needed).
- Upgrade software to the most recent version (please download a copy of your files before sending to LI-COR).
- Check pump and flow rates.
- Other repairs and maintenance as needed; customers are notified if additional fees are required for repairs.
After factory service, the serviced instrument should perform as well as a new instrument.
To request factory service or recalibration for an instrument:
Service will be performed at our US office in Lincoln, NE USA, European office (LI-COR GmbH), or by your local distributor.