Factory Calibration
LI-830, LI-850, and LI-870 recommended recalibration frequency: No regular factory recalibration required.
What's included with factory calibration?
A factory recalibration includes the following:
Calibration
- Analyzer drift and stability is checked.
- H2O analyzer calibration (6 points, 3 temperatures; LI-850 and LI-870 only).
- CO2 analyzer calibration (13 gas concentrations at 3 temperatures).
- The accuracy of the calibration function is then verified across the measurement range of CO2 and H2O (LI-850 and LI-870 only).
Tests and Maintenance
- The optical path components and hoses are inspected.
- Software updated to the most recent version.
- Other repairs and maintenance as needed; you'll be notified if additional repairs are required.
To request factory service or recalibration for an instrument:
Service will be performed at our US office in Lincoln, NE USA, European office (LI-COR GmbH), or by your local distributor.