Software change log
Version 2.3.8 (released March 6, 2023)
- Added High Altitude configuration.
Version 2.3.5 (released December 8, 2022)
- Improved H2O correction algorithm in LI-7820 instruments.
- Updated algorithms to improve drift performance in LI-7820 instruments.
Version 2.3.3 (released October 11, 2022)
- Version 2.3.3 introduces compatibility with a new wireless chipset. No changes to the interface, operation, or performance.
Version 2.1.41 (released February 10, 2022)
New Features
- Added support for port forwarding.
Bugs Fixed
- Convert pump tach sync to more accurate math for tighter control.
- Improve temperature control sensitivity to reduce cavity or thermal enclosure oscillations.
- Fixed laser power measurement for LI-7820 instruments.
- Changed dither cavity pressure set point to improve control loop.
Version 2.1.37 (released September 24, 2021)
- Improved vibration-induced noise performance.
- Removed A, B, C Pump Adjustment setting; no longer needed due to the improved vibration-induced noise performance.
Version 2.1.35 (released August 18, 2021)
- Changed graph decimal precision from five places to three places.
- Changed the way graph tool tips are displayed to ensure correct placement at all screen sizes, including unconventional or uncommon dimensions.
- Improved detection of user browser/mobile device settings to improve experience across platforms.
- Fixed bug where History graph would show incorrect end times at certain zoom levels.
- Fixed bug where certain screen resolutions would render graph tool-tips unreadable.
- Fixed bug where Android phones could enter a Remark of more than 15 characters.
- Fixed bug where failing to include end time on a data export would crash the download.
- Fixed bug where above bug fix would crash download specifically on Safari (all blank end times are now automatically changed to 00:00 of the following day).
- Fixed bug where History graphs would fail to resize in Safari.
- Fixed various responsive styling issues in Safari.
- Fixed bug where Low Flow Config modal was no longer visible on Settings page.
- Fixed bug where header information (serial number, connection status, etc.) was no longer visible in Spanish language mode.
- Fixed bug where SD card logging errors in header would not change to reflect selected language mode.
- Improve second ring-down logic to improve stability when there is a skipped mode.
- Added functionality to improve handling of fast gas concentration transients.
- Fix default phase setpoint for TG20.
- Remove feature to try to keep timestamp at 0 fractional seconds which would break the averager.
- Remove alternation in zero cal and when initializing start frequency.
- Expand TG20 self cal laser temperature range.
- Updated interface so it is compatible with a new browser update.
- Remove N2O spectra weight from TG10 and TG15 log file.