Factory Calibration
LI-7700 Recommended Recalibration Frequency: No regular factory recalibration required.
What is Included with LI-7700 Recalibration?
A factory recalibration includes the following:
Calibration
- Analyzer stability is checked across the operating temperature range.
- CH4 analyzer calibration (7 points, 3 temperatures).
- The accuracy of the calibration function is then verified across the measurement range of CH4.
- Pressure transducer calibration.
- Temperature sensors verification.
Tests and Maintenance
- Clean analyzer, if needed.
- Change detector housing chemicals.
- Checked for leaks and repaired as needed.
- Lithium coin cell battery changed as needed.
- Check pump control, as well as motor and heater.
- Upgrade software to the most recent version.
- Other repairs and maintenance as needed; you'll be notified if additional repairs are required.
To request factory service or recalibration for an instrument:
Service will be performed at our US office in Lincoln, NE USA, European office (LI-COR GmbH), or by your local distributor.