Factory Calibration
Recommended Recalibration Frequency: No regular factory recalibration required.
What's included with LI-710 recalibration?
A factory recalibration includes:
Calibration
- Calibrate sonic anemometer (wind speed at 0 within 0.05 RMS; sonic temperature within ±0.2 °C).
- Calibrate flow meter across range and verify performance.
- Calibrate RH sensors (up to 7 dew points at 3 temperatures).
Tests and Maintenance
- Check and replace filter as needed.
- Check and replace O-rings as needed.
- Update device firmware.
- Verify functionality and performance.
- Other repairs and maintenance as needed. Customer approval is required for additional repairs.
After factory service, the serviced instrument should perform as well as a new instrument.
To request factory service or recalibration for an instrument:
Service will be performed at our US office in Lincoln, NE USA, European office (LI-COR GmbH), or by your local distributor.