Factory Calibration
Recommended Recalibration Frequency: No regular factory recalibration required.
What's included with LI-610 recalibration?
A factory recalibration includes:
Calibration
- Water pump is filled and tested.
- Dew point temperature calibration (2 points).
- Check panel meter accuracy.
Tests and Maintenance
- Test battery.
- Replace internal air filter.
- Clean fan filter and hose barb screen as needed.
- Inspect condenser block.
- Replace hoses and other air supply parts as needed.
- Check and repair any air leaks in the system (inform customer of leaks found).
- Other repairs and maintenance as needed; customers are notified if additional fees are required for repairs.
To request factory service or recalibration for an instrument:
Service will be performed at our US office in Lincoln, NE USA, European office (LI-COR GmbH), or by your local distributor.