Most computers do not provide enough power through USB ports to charge the LI-600 satisfactorily. It may take longer than a full night to charge from a computer. We recommend using the charger provided with the LI-600 or another compatible quick-charger for the fastest charging.
No, it is not necessary to calibrate the sensors before using it for the first time. See the LI-600 Porometer/Fluorometer Quick Start Guide to learn more about taking your first measurement.
This is most likely an issue with one or more calibration parameters in your fluorometer. Therefore, check those parameters first. To do this, connect the LI-600 to the computer software and enable the Factory Calibration in the software (Figure 1‑1). Then go to the factory page (Figure 1‑2) and view the values for the fluorometer. Use the Automatic Factory Reset to update them, but you can also manually enter them, and then press ‘Calibrate’ to load them into the instrument.