Factory Calibration
Recommended Recalibration Frequency: No regular factory recalibration required.
What's included with LI-600/N recalibration?
A factory recalibration includes:
Calibration
- Calibrate flow meter across range (10 flow rates across flow range). Performance is verified after calibration.
- Calibrate RH sensors (dry air and up to 6 dew points at 3 ambient temperatures). Performance is verified after calibration.
- Infrared leaf temperature sensor (calibration is derived from 5 temperature measurements of a simulated leaf in 3 ambient air temperatures that span the operating temperature range). Performance is verified after calibration.
Tests and Maintenance
- Check filters and replace if needed.
- Check gaskets and replace if needed.
- Test battery.
- Verify functionality.
- Other repairs and maintenance as needed. Customer approval is required for additional repairs.
To request factory service or recalibration for an instrument:
Service will be performed at our US office in Lincoln, NE USA, European office (LI-COR GmbH), or by your local distributor.