Specifications
- Power Supply: 110 – 240 VAC; 50 or 60 Hz
- Precision: Better than ± 0.001 pH units
- Accuracy: About ± 0.003 pH units
- Sample Volume: ~ 40 mL
- Time Per Sample: ~ 5 minutes per measurement
- pH Indicator: m-Cresol Purple
- Spectrophotometer: Agilent Cary 60
- Spectrophotometric Flow Cell: 10 cm optic length with a thermostated water jacket
- Operating Environment: Indoor (both land and shipboard laboratories)
- Operating Temperature Range: Normal room temperature (15 – 40 °C)
- Operating Humidity Range: <95% RH (non-condensing conditions)
- Temperature Control: ± 0.1 °C
- Communication: RS-485 and USB (USB-to-RS-485 converter is included)
- Dimensions(LI-5700A Control Unit): 28 cm × 30 cm × 24 cm (H × W × D)
Specifications subject to change without notice.