Specifications
Power Supply Voltage: 5 - 32 VDC
Power: 600 mW (500mW typical)
Operating Environment:
Altitude: Up to 5000 meters
Temperature: -20 to 72 °C
Relative Humidity: 0 to 100% (non-condensing)
The 550USB is not rated for outdoor use.
Transient Overvoltage: Category I
Temporary Overvoltage: 40 V
Pollution Tolerance: Level 3 (enclosure must be intact)
Sampling Frequency: 100 Hz maximum
Ultrasonic Frequency: 60 kHz ±5 kHz
Weight: 225 grams
Size: 10.2 x 10.2 x 24.9 cm
Digital Output: RS-232, RS-422, UART-3V
Output Rate: 1 Hz, 2 Hz, 5 Hz, 10 Hz, 20 Hz, 25 Hz, 50 Hz, 100 Hz
Wind Speed Measurements:
Range: 0-50 m/s
Resolution: 0.01 m/s
Accuracy: (0-10 m/s): ±1%*
Accuracy: (11-30 m/s): ±1%
Accuracy: (31-50 m/s): ±2%
*Averaged over 10 seconds
Wind Direction Measurements:
Range (u/v): 0-359°
Range (w): ±60°
Resolution: 1.0°
Accuracy: ±1.0°
Temperature Measurements:
Range: -20 °C to 72 °C
Resolution: 0.01 °C
Accuracy: ±2.0 °C
Accelerometer:
Range (u, v, w): ±2 g
Tilt (pitch, roll): ±90°
Specifications are subject to change without notice.