Specifications
Resolution: 1 mm2 or 0.1 mm2 (user-selectable).
Scanning Area:
1 mm2 Resolution: 1 mm × 1 mm.
0.1 mm2 Resolution: 0.300 mm L × 0.33 mm W.
Accuracy:
Sample Area | |||||
---|---|---|---|---|---|
Resolution | 10 cm2 | 5 cm2 | 1 cm2 | 0.5 cm2 | 0.25 cm2 |
1 mm2 | ±1% | ±2% | ±5% | ±7% | -- |
0.1 mm2 | ±0.5% | ±1% | ±1% | ±1.5% | ±4% |
Combined accuracy and precision to 99% confidence with correct calibration on verifiable shapes. Accuracy can be increased by calibrating and using in a specific zone of measurement (e.g. the center). Samples with a high proportion of edges, such as conifers, roots, wires, etc. will raise uncertainties to 5% and more. Use only 0.1 mm2 resolution for these types of samples.
Display Capacity:
1 mm2 Resolution: 999,999.99 cm2
0.1 mm2 Resolution: 99,999.999 cm2
Display: 8-digit LED (light emitting diode).
Sample Dimensions:
Width: 25.4 cm maximum; 1.5 to 3 mm minimum.
Thickness: Up to 2 cm, expandable to 2.5 cm by the user.
Length: Unlimited.
Conveyer Belt Speed: 8.0 cm s-1 at 60 Hz; 6.7 cm s-1 at 50 Hz.
Transparent Belts: Rugged clear vinyl.
Light Source: 15 watt fluorescent tube.
Power Requirement: 108-126/216-252 VAC, 48-66 Hz, 100 W max.
Operating Temperature: +15 to 55 °C.
Storage Temperature: -20 to 65 °C.
Size: 25.0 H × 60.0 W × 73.0 cm L (9.8" × 23.6" × 28.7")
Weight: 43 kg (95 lb.).
