Here we describe how to connect sensors and enter calibration information for the basic, standard, and premium sensor packages from LI-COR. Although we show how to connect the Stevens HydraProbes, they must be connected and configured one after the other until each one has a unique address.
Connecting sensor wires for the Basic Package
The basic biomet system supports sensors listed in Table 3‑1.
Wiring table
Table 3‑1. Wiring for the basic biomet system 7900-109. Sensor wires that are not mentioned here are not used and can remain unconnected.
Measurement
Description
Wire Color
Terminal
Soil Heat Flux Plate 1 (Hukseflux HFP01)
Signal +
White
A1
Signal -
Green
A2
Shield
Black
G1
Soil Heat Flux Plate 2 (Hukseflux HFP01)
Signal +
White
A3
Signal -
Green
A4
Shield
Black
G2
Soil Heat Flux Plate 3 (Hukseflux HFP01)
Signal +
White
A5
Signal -
Green
A6
Shield
Black
G3
Soil Moisture Probe 1 (METER or Decagon EC-5)
Signal +
Orange or Red
A7
Ground
Bare
G4
Power
Brown or White
A15EXC
Soil Moisture Probe 2 (METER or Decagon EC-5)
Signal +
Orange or Red
A8
Ground
Bare
G4
Power
Brown or White
A16EXC
Air Temperature and Relative Humidity (Vaisala HMP155)
Temp. Signal
White
A9
RH Signal
Yellow
A10
Shield
Black
G5
Signal Ground
Green
G5
Power Ground
Red
GND
Power
Blue
12V
Net Radiation (NR Lite2)
Signal +
Red
A11
Signal -
Blue
G6
Shield
Black
G6
Soil Temperature Probe 1 (LI-COR 7900-180)
Signal +
Red
A12
Ground
Black
G6
Power +
Orange
A15
Soil Temperature Probe 2 (LI-COR 7900-180)
Signal +
Red
A13
Ground
Black
G7
Power +
Orange
A16
Soil Temperature Probe 3 (LI-COR 7900-180)
Signal +
Red
A14
Ground
Black
G7
Power +
Orange
A16
PPFD (LI-CORLI-190R)
Cathode
Blue
G9
Anode
Brown
A18
Shield
Bare
G9
Global Radiation (LI-CORLI-200R)
Cathode
Blue
G8
Anode
Brown
A17
Shield
Bare
G8
Liquid Precipitation (Texas Electronics TR-5252M)
Switch
Red
C1
Switch
Black
GND
Wiring diagram
Sensor wiring for the basic biomet system (7900-109).
Connecting sensor wires for the Standard Package
The basic biomet package supports sensors listed in Table 3‑2.
Wiring table
Table 3‑2. Wiring for the standard biomet package 7900-118. Sensor wires that are not mentioned here are not used and can remain unconnected.
Measurement
Description
Wire Color
Terminal
Soil Heat Flux Plate 1 (Hukseflux HFP01)
Signal +
White
A1
Signal -
Green
A2
Shield
Black
G1
Soil Heat Flux Plate 2 (Hukseflux HFP01)
Signal +
White
A3
Signal -
Green
A4
Shield
Black
G2
Soil Heat Flux Plate 3 (Hukseflux HFP01)
Signal +
White
A5
Signal -
Green
A6
Shield
Black
G3
Net Radiation Cable 1 (Hukseflux NR01-T2)
Upper Pyranometer +
Red
A7
Upper Pyranometer -
Blue
A8
Ground
Black
G4
Lower Pyranometer +
White
A9
Lower Pyranometer -
Green
A10
Upper Pyrgeometer +
Brown
A11
Upper Pyrgeometer -
Yellow
A12
Lower Pyrgeometer +
Pink
A13
Lower Pyrgeometer -
Gray
A14
Net Radiation Temperature Cable 2 (Hukseflux NR01)
Ground
Black
G6
Thermistor +
Red
A15
Thermistor +
White
A15
Thermistor -
Blue
G7
Thermistor -
Green
G7
Heater Ground
Yellow
GND
Heater Power
Brown
SW1
Not Connected
Pink and Gray
Not Used
Air Temperature and Relative Humidity (Vaisala HMP155)
Temperature Signal
White
A16
RH Signal
Yellow
A17
Shield
Black
G8
Signal Ground
Green
G8
Power Ground
Red
GND
Power
Blue
12V
PPFD (LI-CORLI-190R)
Cathode
Blue
G9
Anode
Brown
A18
Shield
Bare
G9
Liquid Precipitation (Texas Electronics TR-5252M)
Switch
Red
C1
Switch
Black
GND
Soil Moisture and Temperature (Stevens HydraProbe)
Data
Blue
DIN Terminal 8
Ground
Black
DIN Terminal 9
Power
Red
DIN Terminal 10
Wiring diagram
Sensor wiring for the standard biomet package (7900-118).
Connecting sensor wires for the Premium Package
The premium biomet package supports sensors listed in Table 3‑3.
Wiring table
Table 3‑3. Wiring for the premium biomet system 7900-119. Sensor wires that are not mentioned here are not used and can remain unconnected.
Measurement
Description
Color
Terminal
Soil Heat Flux Plate 1 (Hukseflux HFP01)
Signal +
White
A1 (DAqM 1)
Signal -
Green
A2 (DAqM 1)
Shield
Black
G1 (DAqM 1)
Soil Heat Flux 1 Calibration (Hukseflux HFP01SC)
Heater +
Brown
SW1 (DAqM 1)
Heater -
Green
GND (DAqM 1)
Shield
Black
GND (DAqM 1)
Soil Heat Flux Plate 2 (Hukseflux HFP01)
Signal +
White
A3 (DAqM 1)
Signal -
Green
A4 (DAqM 1)
Shield
Black
G2 (DAqM 1)
Soil Heat Flux 2 Calibration (Hukseflux HFP01SC)
Heater +
Brown
SW2 (DAqM 1)
Heater -
Green
GND (DAqM 1)
Shield
Black
GND (DAqM 1)
Soil Heat Flux Plate 3 (Hukseflux HFP01)
Signal +
White
A5 (DAqM 1)
Signal -
Green
A6 (DAqM 1)
Shield
Black
G3 (DAqM 1)
Soil Heat Flux 3 Calibration (Hukseflux HFP01SC)
Heater +
Brown
SW3 (DAqM 1)
Heater -
Green
GND (DAqM 1)
Shield
Black
GND (DAqM 1)
Net Radiometer (Kipp & Zonen CNR4)
Upper Pyranometer +
Red
A7 (DAqM 1)
Upper Pyranometer -
Blue
A8 (DAqM 1)
Lower Pyranometer +
White
A9 (DAqM 1)
Lower Pyranometer -
Black
A10 (DAqM 1)
Upper Pyrgeometer +
Gray
A11 (DAqM 1)
Upper Pyrgeometer -
Yellow
A12 (DAqM 1)
Shield
Bare/Black
G6 (DAqM 1)
Lower Pyrgeometer +
Brown
A13 (DAqM 1)
Lower Pyrgeometer -
Green
A14 (DAqM 1)
Net Radiometer Temperature (Kipp & Zonen CNR4)
Shield
Bare/Black
G7 (DAqM 1)
Thermistor -
Black
G7 (DAqM 1)
Thermistor +
White
A15 (DAqM 1)
PPFD (LI-CORLI-190R)
Cathode
Blue
G9 (DAqM 1)
Anode
Brown
A18 (DAqM 1)
Shield
Bare
G9 (DAqM 1)
Soil Moisture and Temperature (Stevens HydraProbe)
Signal
Blue
DIN Terminal 8
Ground
Black
DIN Terminal 9
Power
Red
DIN Terminal 10
Liquid Precipitation (Texas Electronics TR-525M)
Switch
Red
C1 (DAqM 1)
Switch
Black
GND (DAqM 1)
Air Temperature and Relative Humidity (Vaisala HMP155)
Temperature Signal
White
A16 (DAqM 1)
RH Signal
Yellow
A17 (DAqM 1)
Shield
Black
G8 (DAqM 1)
Signal Ground
Green
G8 (DAqM 1)
Power Ground
Red
GND (DAqM 1)
Power
Blue
12V (DAqM 1)
Net Radiometer Heater and Ventilator (Kipp & Zonen CNF4)
Heater (-)
Blue
GND (DAqM 2)
Heater (+)
Black
SW1 (DAqM 2)
Heater (+)
White
SW1 (DAqM 2)
Heater (-)
Red
GND (DAqM 2)
Ventilator (+)
Yellow
SW2 (DAqM 2)
Ventilator (-)
Brown
GND (DAqM 2)
Shield
Bare/Black
GND (DAqM 2)
Tacho (not used)
Gray/Green
Not Used
Soil Temperature Probe 1 (LI-COR 7900-180)
Signal +
Red
A13 (DAqM 2)
Ground
Black
G6 (DAqM 2)
Power +
Orange
A15 (DAqM 2)
Soil Temperature Probe 2 (LI-COR 7900-180)
Signal +
Red
A14 (DAqM 2)
Ground
Black
G7 (DAqM 2)
Power +
Orange
A15 (DAqM 2)
Soil Temperature Probe 3 (LI-COR 7900-180)
Signal +
Red
A16 (DAqM 2)
Ground
Black
G8 (DAqM 2)
Power +
Orange
A15 (DAqM 2)
Wiring diagram
Sensor wiring for the premium biomet package (7900-119).