Environmental
Environmental
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||
There are many reasons why LI-COR light measurement instruments are utilized worldwide. They have been designed and field proven to meet research demands in agriculture, oceanography and limnology, solar energy, industry, and the biological sciences. LI-COR radiation sensors have been an industry standard for more than 30 years, with more than 100,000 light sensors sold. Experience, attention to detail, and an advanced knowledge of optical science are what separate LI-COR sensors from other alternatives.
Cable TerminationThere are several options available for cable lead termination on LI-COR radiation sensors. The two letters at the end of the model number denote the type of termination (e.g. “SA”). The following table describes these options:
Milivolt AdaptersMillivolt adapters are used only for connecting sensors to a user’s datalogger or stripchart recorder, not the LI-1400 or other LI-COR readout devices. The millivolt adapter connects to the BNC connector of the sensor, and the wire leads of the adapter are connected to the data logger. Sensor output (in millivolts) when using the millivolt adapter can be computed using "Ohm's Law" (Voltage = Current X Resistance). Example: Calculate the millivolt output of an LI-190 Quantum Sensor which has a calibration constant of 8.0 µA/1000 µmol s-1 m-2. Assume the 2290 millivolt adapter is used with the sensor.
Type "SL" sensors are available for the LI-190, LI-191, LI-200, and LI-210 Sensors. The Type SL Sensor produces a standardized millivolt output, and may be used in place of the Type SA Sensor and Millivolt Adapter. 2290 Millivolt Adapter 2220 Millivolt Adapter 2291 Millivolt Adapter Sensor (1210 Ohm resistance).
Calibration of LI-COR Radiation SensorsRecalibration of LI-COR radiation sensors is recommended every two years. Sensors must be returned to LI-COR for recalibration. Please contact LI-COR for a return materials authorization (RMA) number prior to returning any instruments. Importance of Calibration There are two methods by which the readout device used with a sensor may be adjusted to the calibration constant to provide a direct readout. The constant is numerically entered into the instrument's internal electronic memory via the instrument's keypad. Adjustment of the internal "trimpots" (electromechanical memory) is done. Instruments without any provision for adjustment must have their data scaled by a factor determined from the calibration constant. The characteristics of the optical components change with time and may also be affected by environmental conditions. Periodic recalibration is necessary to restore correct measurements. The time intervals suggested by LI-COR for recalibrations should be seriously considered. Data of a critical nature requires more frequent calibrations than those suggested, as well as calibrating under the conditions of actual use (e.g. same temperature). Ultimately, the only way to verify performance is to recalibrate. Basis of Calibration CalconstantsRelationship between the Calibration Constant
and the Calibration Multiplier LI-COR Light Meters and dataloggers measure the current output of the sensor in units of microamps, and convert the measured current to units of radiation. To make this conversion, LI-COR instruments use the sensor Calibration Multiplier. The Calibration Multiplier is the negative reciprocal of the Calconstant. Multiplier = -1/Calconstant The Calibration Multiplier is always a negative number (because the shield of the coaxial cable of the sensor is positive instead of negative), and is expressed in radiation units per microamp. For complete information on configuring your instrument, please consult your instruction manual or contact LI-COR. If the calibration constant for your sensor has been lost or misplaced, it can be obtained from LI-COR by providing the serial number of the sensor. |
LI-250A Light Meter
LI-1400 DataLogger
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||








